Question
How to measure Inline and Crossline Profiles with BP Compact and rotating table?
Answer:
Step 1
Set up the device in the bunker, check the SSD and Isocenter position of the detector.
Step 2
In myQA Accept, set the Device turn angle and create the queue item for one profile direction.
Start measurement.
Step 3
Rotate the phantom with the rotating table. Re-check the Isocenter position of the detector.
Step 4
In myQA Accept, open a new project - check the Device turn angel and create queue item for another profile direction.
Start measurement again.
You can find a video in the attachment below.