Question
How can I turn off susceptibility in the MRID3D application?
Answer
- Close the MRID3D software
- Open a File Explorer instance and navigate to %appdata%\Modus Medical Devices\MRID3D
- Edit the MRID3DEngine.config file in the folder. Turn on the UI by changing the following entries from false to true
- ShowEngineSelector
- ShowSusceptibility
- Optionally set UseSusceptibility to false at this time though this setting can be changed from the Miscellaneous Settings section of the Settings view now.
- Save the edit and close the config file
- Start the MRID3D software again and you will have the Engine Version selection control at the top right of the data grid in addition to the Susceptibility setting in the Settings view NOTE: In the Engine Version drop down tagged with ‘_NC’ at the end is the one with No Susceptibility Correction performed. Analysis will be done according the current Susceptibility setting in the Settings view when it is started and NOT the state of the Engine Version selector.