Question
How can I update the DOSE-X firmware?
Answer
Verify the checksum
Before the installation of a new firmware version, verify the checksum of the downloaded file with the checksum provided by IBA Dosimetry along with the firmware version, to ensure the authenticity and integrity of the received firmware file. Contact your IT Administrator for more information on verifying checksums or the support of IBA Dosimetry.
Updating the firmware via USB flash drive
Please note, the FW can only be updated via USB flash drive when using the touchscreen of the DOSE-X.
If you try to update the FW without a USB flash drive plugged into the device, a dialog will appear prompting the user to insert a USB flash drive.
Please note, that the USB flash drive needs to be formatted in FAT32 and only USB3.0 drives should be used.
Please also note, that for the update, Admin access is required.
Step 1
Upload the FW onto a secure USB flash drive
Step 2
Insert the USB flash drive the into the DOSE-X.
Step 3
Go the Update Firmware page in System Settings.
Step 4
Open the USB flash drive and navigate to the folder containing the latest FW.
Step 5
Tap the UPLOAD button and confirm the installation by tapping YES.