How to use Accu‐Gold 2 on a Siemens CT with Straton Tube
Updated
In order to make seamless measurements on a Siemens CT that uses a Straton tube, a Model 8155 Tungsten Filter has been developed to be used together with a special calibration.
Included on this flash drive is the calibration file needed for using it with the Accu‐Gold software.
Open the AG2 software then choose on the menu to import the license:
Point to the “2018‐09‐18 CORR Somaton‐stratonW.aglicenses” file on the flash drive and click Open.
When making measurements on the machine, select the Anode/Filter combination for Siemens, then Stranton Tube with W filter:
Make sure the sensor has the Tungsten filter in place before making the measurements.
In order to use the appropriate cal file for the Touch, Touch WiFi and the Accu‐Gold Nugget, use the links on our webpage to reinstall the firmware. The firmware will include the cal file automatically.
Got to www.Radcal.com, choose the Support page then Accu‐Gold Software. Download the appropriate firmware and follow the instructions.