For use with Accu-Gold, Accu-Pro, and Accu-Dose, the 10X6-6 is a general purpose in beam chamber with a well-documented wide dynamic range chamber with many dose and rate applications. It is also recommended for measuring exposure time in Auto Dose mode. The range of time (pulse width) extends from 10ms to 300s.
| Chamber | 10X6-6 | |
| Min Rate | 2 µR/s | 20 nGy/s |
| Max Rate | 17 R/s | 149 mGy/s |
| Min Dose | 10 µR | 100 nGy |
| Max Dose | 59 kR | 516 Gy |
| Cine Specifications | 516 Gy | 1 nGy/f - >10 mGy/f |
| Calibration Accuracy | ±4% using X-rays @ 60kVp and 2.8 mm AL HVL | |
| Exposure Rate Dependence | ±5%, 0.4 mR/s to 80 R/s, up to 500 R/s for 50 us pulses | |
| Energy Dependence | ±5%, 30 keV to 1.33 MeV (with build-up material) | |
| Construction | Polycarbonate walls and electrode conductive graphite interior coating; 6 cm3 active volume; 0.05kg | |