For use with Accu-Gold, Accu-Pro, and Accu-Dose, the 10X6-180 is a leakage and low level measurements chamber. The chamber is designed for leakage measurements with a cross-section of 100 cm² (complies with CFR21) and volume of 180 cm³. It is also used for very low dose to image receptors.
| Chamber | 10X6-180 | |
| Min Rate | 100 nR/s | 1 nGy/s |
| Max Rate | 0.6 R/s | 4.9 mGy/s |
| Min Dose | 200 nR | 2 nGy |
| Max Dose | 2.0 kR | 17 Gy |
| Cine Specifications | N/A | |
| Calibration Accuracy | ±4% using X-rays @ 150kVp and 10.2 mm AL HVL | |
| Exposure Rate Dependence | ±5%, 20 mR/hr to 2000 R/hr | |
| Energy Dependence | ±5%, 30 keV to 1.33 MeV (with buildup material) | |
| Construction | Polycarbonate walls and electrode; conductive graphite exterior coating; 180 cm³ active volume; 0.11kg | |