A single sensor leakage measurement solution for very low-level radiation measurements such as shielding, leakage, irradiators and environmental that is part of the Accu-Gold family of systems. Equivalent to the Fluke® 96010A Ion Chamber used in the Fluke® Radiation Leakage Detection System, this chamber meets 21 CFR 1020.30(k) for leakage measurement requirements.
| Chamber | 10X6-500 | |
| Min Rate | 62 nR/s | 0.6 nGy/s |
| Max Rate | 130 mR/s | 1.2 mGy/s |
| Min Dose | 1 nR | 0.01 nGy |
| Max Dose | 45 R | 0.4 Gy |
| Cine Specifications | N/A | |
| Calibration Accuracy | ±4% using X-rays @ 100kVp and 3.96 mm AL HVL | |
| Exposure Rate Dependence | <5% up to 5 Gy/hr (570 R/hr) | |
| Energy Dependence | ±5%, 40-150 kVp (RQR and RQA Beam Qualities) | |
| Construction | Truncated cylinder, polycarbonate walls and electrode; conductive graphite exterior coating; 522 cm3 active volume; 0.54 kg. | |