The 0.6cc thimble ionization chamber is ideal for dose measurements in modern wide beam multi-slice CT. It is calibrated using X-rays @ 150 kVp. Phantom adapter is included.
| Chamber | 10X6-0.6CT | |
| Min Rate | 20 µR/s | 200 nGy/s |
| Max Rate | 133 R/s | 1.17 Gy/s |
| Min Dose | 100 µR | 1 µGy |
| Max Dose | 589 kR | 5 kGy |
| Cine Specifications | N/A | |
| Calibration Accuracy | ±4% using X-rays @ 150 kVp and 10.2mm Al HVL | |
| Exposure Rate Dependence | ±2%, 10 mR/s to 100 R/s | |
| Energy Dependence | ±5% 3 to 20mm Al HVL | |
| Active Length | 19.7 mm ± 1 mm | |
| Construction | C552 air-equivalent material walls and electrode; polyacetal exterior cap, 0.6 cm³ active volume, 3-m triax cable | |