For use with Accu-Gold, Accu-Pro, and Accu-Dose, the DDX6-WL is a solid-state dose sensor for diagnostic range measurements. It is a small solid-state sensor for single exposures and fluoroscopy in the Diagnostic range. It measures dose, dose-rate, and time and is passively corrected for beam quality. It may be used simultaneously with an AGMS-D multi-sensor in entrance / exit exposure measurements. It is especially recommended for low dose-rate measurements, e.g. pediatric dose-rates.
| Sensor | DDX6-WL | |
| Min Rate | 50 nR/s | 500 pGy/s |
| Max Rate | 0.65 R/s | 5.7 mGy/s |
| Min Dose | 200 nR | 2 nGy |
| Max Dose | 1.8 kR | 15 Gy |
| Accuracy | ±5% using X-rays @ 80kVp with 2.5 mm Al total filtration (IEC 61267 RQR-6) | |
| Energy Dependence | +5% from 50 kVp to 120 kVp at 2.5 mm Al | |
| Energy Dependence | +5% to -10% from 2.5 mm Al to 23 mm Al | |