This manual does not cover software and device driver installation. For information on installing the software and driver, see the readme file located on the flash drive included with your Accu-Gold device.
Install Accu-Gold software version 2.
Snap the Nugget onto the digitizer module:
Align the USB connectors and slide the Nugget all the way in. (To remove it, pull out the connector first then unsnap it by tilting it upward or sliding it off the end of the digitizer.)
Plug in a multi-sensor, ion chamber or dose diode.
Press the power button on the Nugget. The power button will turn green. (If the charger is plugged in, it will blink yellow.) The Wi-Fi status light will turn on displaying green and then orange. When it blinks green, it is ready to connect.
Start the Accu-Gold 2 software.
Click on Connect via Wi-Fi. The software will detect the presence of the Nugget and connect to it automatically.
You are now ready to start.
Connecting manually
If the Nugget is not able to connect automatically, you will need to connect manually. If there are multiple Nuggets that show up, you will also have to connect manually.
For Windows 10, if you get this dialog box, click on “Connect using a security key instead ”
On the computer, check the Wireless Network Connection pane.
There, you will find “Nugget_51xxxx” as one of the Wi-Fi devices (the ID number will be the Nugget_serial number). Click on the Nugget and when asked for the password, enter 12345678.
NOTE: Internet Protocol Version 4 (IPv4) must not be set to a fixed IP address. IPv4 must be set to “Obtain an IP address automatically” and “Obtain DNS server address automatically. Contact us if any issues arise.
Start the Accu-Gold 2 software.
Click on “Connect via Wi-Fi”.
You are now ready to start.
Note that not all sensors, measurements and modes of the Accu-Gold software are available when using the Nugget. See Appendix A for a comparison list.