To access system information and change device and measurement settings, tap the Menu button and select System settings.
Please note that some items and actions are only available to users with Admin access.
| Item | Description |
| Device info | Shows device information including, device names, serial number, production date, and hardware and FW versions. Please note, this information is not editable. |
| Display |
|
| Audio |
|
| Diagnostic functions |
|
| Export log files |
Select a log type and tap the Export button (see Section 3.2.2). Available log types:
|
System Settings for users with Admin access only
| Item | Description |
| Time & date |
Click SET to save the changes. |
| Language | Select the display language. The FW currently supports English (US), German, Spanish, French, Italian, Portuguese, Russian, Japanese, Korean, and Chinese. |
| Units |
Click RESET TO DEFAULT button to reset the units. Before starting a measurement, ensure the kTp correction units are correct. |
| Trigger Mode |
Set the desired values for Trigger Mode (see Trigger mode settings):
|
| User calibration |
Click on High, Medium, or Low to enter the desired sensitivity. Sensitivity ranges:
User calibration factors applied can be switched on/off. |
| Network configuration |
Connect the DOSE-X to a network (WLAN or Ethernet)
See Section 3.3. |
| Admin settings |
|
| Update firmware | Update firmware via USB flash drive or remote access. See Section 6.2.1. |
| Configuration backup |
See Section 3.4. |
Diagnostic Functions
Self-Test Functionalities
Self-tests are performed automatically during device warm-up and can be performed manually at any time by pressing the Perform Test button on the Diagnostic Functions page.
When manually performing self-tests, please ensure that the protective cover has been placed over the Triaxial Connector before tapping the Perform Tests button.
The Self-Test Functions tab displays the results of the Leakage, Bias, and Electrical Sensitivity tests.
- Bias Test: Checks the insulation resistance between the center electrode and GND of the entire sensor, extension cable, and dosimeter system.
- Leakage Test (Isolation Collection electrode to Guard): Checks the insulation resistance between the
center electrode and the guard of the entire sensor, extension cable, and dosimeter system. - Electrical Stability Test: During the test, a highly stable, bipolar reference charge source is used to check the proper function and the long-term stability of the charge input amplifier circuit. The results of these tests are displayed, as absolute values (nominal and measured value) and relative deviation (in percent of nominal value).
System Monitoring
The System Monitoring tab displays the following system status information:
- Power Supply Voltage
- Real Time Clock Battery
- Internal System Voltages
- System Temperature
Export Log Files
Tapping EXPORT LOG FILES in the system settings menu to open the Export log files screen. Select one of four options to export:
- System log: Contains log entries of all FW components, inclusive Kernel. The complete system log may have many entries, resulting in a big export file size. The export file size may be reduced by selecting one of the other three options.
- System log since last boot: Same as above but reduced to the log entries since the last power on/off.
- Measurement log: A filtered system log which contains the log entries from the measurement components only.
- Measurement log since last boot: Same as above but reduced to the log entries since the last power on/off.
DOSE-X data is exported as plain text files which can easily be opened in most text editors and word
processors.