User calibration factors will be applied to measured current / dose rate when displaying the measurement result on the screen.
- Tap User Calibration in the System Settings menu to open the User Calibration page.
- Select the High, Medium, or Low sensitivity range to modify the calibration factor
- Tap ENTER to confirm the new calibration factor.
CAUTION: DOUBLE CHECK USER CALIBRATION FACTORS
Before making changes to the calibration factors, please ensure the value entered in the library is identical to the one which is stated on the ADCL calibration certificate. Double-check detector type and serial number. Incorrect values will put the device out of calibraiton.
For information on Electrical and Dosimetric Calibration, please see Section 6.1 for more information.