| Description | Recommended Action |
|
Internal power supply error The supply voltage provided by the internal power-supply to the DOSE-X electronics is out of the allowed range. |
Power-cycle the device (turn it off, disconnect If the error persists, contact IBA Dosimetry |
|
Internal system voltage out of range. An internal system voltage has been detected to be out of range. |
|
|
Checksum error in non volatile Baseboard memory Memory corruption has been detected. |
|
|
Checksum error in non volatile Electrometer memory Memory corruption has been detected. |
|
|
System fan malfunction The system fan does no longer rotates, turn off the DOSE-X to avoid overheating! |
|
|
Internal system voltage out of range An internal system voltage has been detected to be out of range. |
Turn off the device and wait for 5 minutes before turning it back on. If the error persists, contact IBA Dosimetry Service. |
| Admin authentificaiton failed |
Ensure that the password is correct. In the event of a lost admin password, user |
| Admin password has been reset | Contact your system admin for the new admin password. |
|
Detector Bias Voltage out of range. The detector bias voltage could not be to the required level, this may be caused
|
Check if the detector guard (when grounded), or Bias Voltage input is properly isolated, not shorted, and that the detector is suitable for use with bias (and no coax BNC adapter is used). |
| Detector k_Tp values out of range | Please check the entered values for k_Tp. |
| System temperature out of range |
Only operate the DOSE-X within the given environmental range conditions (see Section 7.8). If the local conditions have been verified to fall within the correct range, a fault of the internal sensor is likely. Contact IBA Dosimetry Service. |
| System humidity out of range |
|
| System air pressure out of range |
|
|
Error in data from Analog to Digital Converter In the event of a surge phenomenon (crosstalk caused by switching very large loads on the DOSE-X supply cable/power grid or a lightning strike in the power grid), the current measurement is aborted for safety reasons. |
Before restarting the measurement, check the If there are any issues with the device after |
|
Overcurrent on electrometer input The current into the electrometer input exceeds the maximum rated input current for the range selected. |
Select a range with lower sensitivity. If the problem persists, and it has been verified that the input current does NOT exceed the value specified in the Instructions for Use, a defect of the electrometer input is possible, contact IBA Dosimetry Service. |
|
Power-on self tests failed The power-on self-test may fail if an external detector is connected. |
To verify correct operation, place the protective cover on the electrometer input connector, then start the self-test in the menu. If the message persists, even with the protective cover placed on the electrometer input, contact IBA Dosimetry Service. |
|
Real time clock battery voltage low The voltage of the internal real-time clock battery is low and probably needs to be re-charged. |
Leave the DOSE-X turned on for a few hours. If the message persists, contact IBA Dosimetry |
| An empty dose rate name is not allowed |
The Dose rate field in the Detector Library |
| Dose rate name already in use |
Each Dose rate entered into the Detector |
| Empty k_Q correction name is not allowed |
Each k_Q correction in the Detector Library must have a name (i.e., entered value). Please ensure that both the Energy type and kQ |
| k_Q correction name is already in use |
Each k_Q entered into the Detector Library must have a unique name (i.e., entered value). |
| An empty library name is not allowed |
Each Detector or Machine in their respective |
| Library name already in use |
Each Detector or Machine in their respective |