The latest FW can be obtained from IBA Dosimetry, please see contact details in Chapter 8.
CAUTION: DO NOT USE OLD FIRMWARE
The installation of an older firmware version may reduce the security or safety of the device.
CAUTION: VERIFY THE CHECKSUM
Before the installation of a new firmware version, verify the checksum of the downloaded file with the checksum for this file provided by IBA Dosimetry along with the firmware version, to ensure the authenticity and integrity of the received firmware file. Contact your IT Administrator for more information on verifying checksums or the support of IBA Dosimetry (Chapter 8).
Updating via USB flash drive
Please note, the FW can only be updated via flash drive when using the touchscreen of the DOSE-X.
If the user tries to update the FW without a flash drive plugged into the device, a dialog will appear
prompting the user to insert a USB.
- Upload the FW onto a secure flash drive.
- Insert the flash drive into the DOSE-X.
- Go to the Update Firmware page in System Settings.
- Open the flash drive and navigate to the folder containing the latest FW.
- Tap the UPLOAD button.
- Confirm the installation by tapping YES.
Updating via Web Browser
- Save the FW on your local PC.
- Open a web browser and access the DOSE-X remotely.
- Go to the Update Firmware page in System Settings.
- Click the BROWSE button and select the correct FW file.
- Click the UPLOAD button
- Confirm the installation by clicking YES
If the DOSE-X is operated remotely, the browser will open a file dialog, allowing the user to select a file from the remote client.
CAUTION: USE THE CORRECT WEB BROWSER
Accessing the firmware UI with a web browser other than specified may introduce risks related to the product safety, usability, or cybersecurity.