Bluetooth FW Upgrade
- Open command shell as administrator.
- Execute “cd C:\Program Files (x86)\IBA Dosimetry\myQA SRS Service \BluetoothFirmwareUpdate” (or the selected installation path during setup)
-
Execute IBADos.Core.UMS.SBRT.BluetoothFirmwareUpdate.exe
Expected output:
-
Now the Bluetooth module is in the state to be updated. The update process is done by a mobile device with a specific App.
IBA Dosimetry will provide the *.gbl file. If more information is needed, please visit the following website for detailed information or contact the IBA Customer Support (see contact information in Chapter 10).
- After that, follow the instruction “Afterwards reboot the computer or restart the service ‘IBADos.Core.UMS.SBRT.Server.exe’ manually”
FPGA Firmware Upgrade
The FW for the FPGA (Field Programmable Gate Array) used in the myQA SRS may subject to be updated over time. IBA Dosimetry Service will inform you the update and the new FW update bin file can be downloaded in the IBA service Portal.
- Download the latest version of FW update via IBA Service Portal: https://iba-portal.com (also see the ticket received for the exact link). Store it in your desired location.
-
In an administrator command shell, execute
IBADos.Core.UMS.SBRT.FpgaFirmwareUpdate.exe with the bin file containing the new FPGA FW as argument. See an example below:If myQA SRS Service Setup is installed in the C drive (see the previous section), IBADos.Core.UMS.SBRT.FpgaFirmwareUpdate.exe is located in C:\Program Files (x86)\IBA Dosimetry\myQA SRS Service\FpgaFirmwareUpdate. See an example below; the FW update file is stored in the same directory as the *.exe file.
- Example of the expected output:
- Switch the device off and on again (as stated in the output).