Question
When measuring with the myQA SRS Cyberknife and the plan path xml file, for some plans I receive the following error message:
The detected radiation times do not correlate to the number of MUs in the plan path xml file.
What can I do to solve this?
Answer
Please increase the frame rate by lowering the "Sample time", e.g. from 0.5 s to 0.2 s.