Subject: Inconsistent 1D gamma analysis results
Products affected: myQA Accept 9.1 and 9.2
Function affected: 1D gamma analysis results
We have identified an issue with the 1D gamma analysis results in myQA Accept (9.1 and 9.2).
When calculating 1-D Gamma Results, the SW will always use the default parameters for the first calculation. This calculation of default values occurs even if the parameters are changed before running the first calculation.
Recommendation:
To avoid confusion, it is advised to first calculate the 1-D Gamma results based on the default parameters. Once the first calculation has been performed, the user can then update the parameters and run additional calculations.