Question
How do I perform a uniformity calibration w/ LDA 99SC detector?
Answer
Step 1
Place the LDA-99 SC in the --/++diagonal position and move it as low as possible (e.g., 30 cm) to have a larger field size.
Step 2
Open the field size as far as possible (normally 40 cm × 40 cm).
Step 3
In Queue Setup > Queue Settings, set the Detector Orientation to the --/++ diagonal position and the Device turn angle to 270˚.
Step 4
Click the Measurement button in the Tasks panel
Step 5
In Controller panel, click Connect, and then perform Gantry coord., Limits, Background, Auto gain, Normalization.
Step 6
In Electrometer panel, select the Calibration tab and then enter the values in the Dose rate and Sampling time boxes.
Step 7
enter the energy used for calibration in the text field.
Step 8
Turn on the beam and then click the Calibrate button.
Step 9
Switch off the beam and click OK in the Background measurement popup message to start the background measurement.