Question
How to upgrade firmware for CCU and hand control?
Answer
Note: if the CCU firmware version is earlier than 3.24, you need to update the CCU firmware to version 3.24 or later first. Then the remote control software can be updated.
Note: if the firmware version of your hand control is earlier than 2010, you cannot upgrade yourself. If you want the upgrade, please contact IBA Dosimetry Service
Please follow the below steps to update firmware:
Step 1
Close all programs that may connect to CCU (eg. myQA Accept, XXsetup)
Step 2
Connect CCU to your workstation, power on CCU.
Step 3
Run XXsetup, click Yes on below window
Step 4
- Update CCU firmware
Press Update firmware... button, open and browse the firmware file, and click open. Software will ask your confirmation on firmware update, click Yes.
CCU will reboot and the new firmware will be enabled
- Update hand control firmware
Press Update remote control firmware... button, open and browse the firmware file, and click open. Software will ask your confirmation on firmware update, click Yes.
Hand control will reboot and the new firmware will be enabled.