Question
How to create PDD and TMR tables in myQA Accept?
Answer
Step 1
Select and open your required PDD or TMR scans
Step 2
Select show data in tabular form
Step 3
Select the table type.
- Single Scan View (one table per scan incl. x/y/z coordinates ) and
- Multiple Scan View (one table showing all scans with only relevant coordinates)
Step 4
For Single Scan View
- Select the Display of data (All or selected) and Point distance.
- Then click on Create Table.
Step 5
For Multi Scan View
- Assuming there is no template to convert the data available, first select the correct Template type. (It will be Depth vs. Fieldsize for PDD scans and TPR/TMR for TMR scans) .
- click on Edit/Add Template(s).
Step 6
In the calculation Details window select the desired resolution of the data table and Renormalization (normally No normalization is selected).
Step 7
Select the Square field size to convert from Single or Multiple options.
Single option:
- Enter the smallest field size in Start box and click on Add.
- Enter the largest field size to be converted and click on Add.
- Then check the box for 'Include source data'. This way all the field sized that are available in PDD data will be converted to table.
- If any interpolated field size is required in the tabular data that can also be added in the Start box.
Multiple option:
- Enter the smallest field size in Start box, largest in Stop box and add the field size increment from smallest field size in the Increment box.
Step 8
Go back to Step 5. In the popped up window and select the template. Select All data and click on Create Table.
Step 9
Right Click on the table-->Copy Table-->Paste into Excel.