Question
How to set limits for detector movement?
Answer
By default, the movement range of the detector is the maximum dimension of the phantom. All limits can be seen in the Position panel of the Controller in the myQA Accept software.
However, if a large detector is used, the maximum range may cause the detector to crash into the wall of the phantom. The sub-limits are used for setting sub-limits for the necessary dimensions.
Step 1
In the Water phantom main display, press the softkey to open the Sub-limit menu:
To set a sub-limit, move the detector to the desired position, press the corresponding softkey. When a green checkbox appears on the right-side of the icon, the sub-limit is set.
To reset the limit to the maximum range, press the corresponding softkey until the check mark disappears (about 2 seconds).