I install Omnipro Advance on a new workstation, how to restore Omnipro Advance database?
Answer
There are two databases in Omnipro Advance, one is the Configuration database, the other is Archive database. The configuration database contains the equipment, the created test queues, and the associated reference measurements. The archive database contains the performed tests with all the associated test parameters (field size, energy, …). It also contains tolerances and measured values of the tests.
Please backup both databases before working on them.
Restore Archive database
Run Omnipro Advance - Database - Open, select the Archive database, and click open.
Restore Configuration database
Step 1
Copy the configuration.fdb to the new workstation on C:\ProgramData\IBADosimetry\OmniPro-Advance\1.2.11.0\Database
Please note, 1.2.11.0 is the version number of Omnipro Advance. When you have a different version, the number here is also different.
Step 2
Run Omnipro Advance again, the previous configuration should be restored already.