Question
In the CBCT plugin test setup, I select a different slice and hit the process button, but the analysis results stay the same. How do I set reference slices in myQA Machines CBCT plugin correctly?
Answer
In order to set reference slice in myQA Machines CBCT plugin, please follow below steps:
Step 1
Click Import folder to import CT slices
Step 2
Select Phantom model, here we take Phantom model Catphan 503 as an example. There 3 different modules in this phantom model, we need to select reference slices one by one.
Step 3
Click the first module (CTP404). In the Image window, you are able to scroll up/down the slice bar to view each slice. When you find the reference slice, click the button 'Set as reference'. The button will then show you 'Is reference', that means this slice has been set as reference.
Step 4
Repeat Step 3 for the other modules.
Step 5
Click Process to get the baseline values. Click OK to save your settings.