Pre-filled Calibration Requests (Medical Imaging)
DosimaxPlusA_RQA_Standard_SSDLCalibration.pdf
This calibration request should be used for the standard calibration of an IBA RQA sensor together with a Dosimax Plus A. It is advisable to choose this calibration for your devices if you conduct a acceptance test of your x-Ray machine.
DosimaxPlusAHV_DCT10_Standard_RQR_SSDLCalibration.pdf
This calibration request should be used for the standard calibration of an IBA DCT10-RS sensor together with a Dosimax Plus A HV. It is advisable to choose this calibration for your devices if you conduct a constancy check of your CT scanner with a filtering of 5.0mm Al using 120kV.
DosimaxPlusAHV_DCT10_Standard_RQT_SSDLCalibration.pdf
This calibration request should be used for the standard calibration of an IBA DCT10-RS chamber together with a Dosimax Plus A HV. It is advisable to choose this calibration for your devices if you conduct a constancy check of your CT scanner with a filtering of 8.5mm Al using 120kV.
MagicMaX_DCT10-MM_Standard_SSDLCalibration.pdf
This calibration request should be used for the standard calibration of an IBA DCT10-MM chamber together with a MagicMaX. It is advisable to choose this calibration for your devices if you conduct a constancy check of your CT scanner with a filtering of 5.0mm Al using 120kV.
MagicMaX_XR_Standard_SSDLCalibration.pdf
This calibration request should be used for the standard calibration of an IBA XR sensor together with a MagicMaX. It is advisable to choose this calibration for your devices if you conduct a constancy check of your X-Ray machine.
Choose RQR5 for checks using no filter with 70kV.
Choose RQA5 for checks using a filter with 70kV.
However, it is always best to choose both qualities for a SSDL calibration.