Question
Is it possible to use the LINAC's or Proton Therapy system's monitor chamber as a reference detector in CCU?
Answer
It is best to know the output current of the monitor chambers to be connected. CCU can only have a maximum of 4 μA. Any detector which has an output current above this value will damage the CCU.