Question
I am performing scans in myQA Accept. Now I see the error "NaN" in the positioning panel and cannot continue my scans. What should I do?
Answer
-
Access Common Settings
- Navigate to the Common Settings section in myQA Accept.
-
Go to Detectors
- Select the Detectors option.
-
Check EPOM Correction
- Ensure the EPOM correction is enabled for photons and electrons (and protons if applicable).
- Verify that a value is set for the EPOM. If you do not wish to use an EPOM correction, enter 0 in the field.
- It is crucial to set an EPOM correction (even if it is zero) once the Depth correction is enabled.
-
Return to myQA Accept
- Go back to myQA Accept, select the chambers with the EPOM correction, and start measuring.