MagicMaX Measurement Software Version 2.2.10
- Improvement for User Convenience
- Excel templates can be installed with the setup.
- Minimum app window size has been reduced.
- New measurement setting "Keep History" has been added, which allows to keep only the last measurement in the session or all measurements.
- PPV result is calculated even if the kV signal is above the limit
- "Dose/Pulse" result is displayed in fluoro mode even with only one pulse
- German PTB approved devices:
Measurement is not considered "legal" if the dose and/or dose rate is below a defined threshold. - German PTB approved devices:
Measurement results are only marked as "legal" at the end of the recording - Improved waves rendering for short measurements with low dose rate.
- A minimum measurement duration of 1ms is applied for the calculation for measurements that are shorter. This is applied for all sensors.
- For measurements without a kV signal, the configured pre- and post-inclusion interval is included in the wave.
- Dose Inclusion configuration is applied to dose calculation for all sensors
- Changed System requirements
- Microsoft Windows 7 SP1 or higher
- Microsoft .NET Framework 4.8
MagicMaX Measurement Software Version 2.2.9
- Improvement for User Convenience
- Invasive current probe measurements are now possible with the current probe as a trigger, eliminating the need for a sensor in the beam.
-
Users can use the Stop delay to specify how long the MagicMaX waits after the trigger threshold is undershot until a measurement is considered complete.
For example, use a longer stop delay when measuring with a very low frequency pulsed fluoroscopy system or to include a pre-pulse in the same measurement as the following “normal” exposure. - Correction factors for custom temperature or humidity settings are always visible at the top of the measurement screen for the CT chamber.
- To prevent inaccurate readings, kV measurement results are hidden and a warning message is displayed when users measure outside of the total filtration specification.
MagicMaX Measurement Software Version 2.2.8
- This version was primarily developed to meet the requirements of the German PTB for the Ionization Chamber 10XF-3CT release process.
- Improvement for User Convenience
- Enhanced the export function by updating the column sorting when exporting data to Excel.
MagicMaX Measurement Software Version 2.2.7
- Improvement vs Version 2.2.6
- Sensor setup dialog always opens even if only one sensor is configured.
- "Auto Cal" function is set as fixed default for XM sensors in sensor setup.
This function only applies to MagicMaX Universal models approved by German authorities (Eichamt).
MagicMaX Measurement Software Version 2.2.6
- Changed System requirements
- Removed Windows XP and Windows Vista Support
- Removed Microsoft Excel 2003 and 2007 support
- Added Windows 10 and Microsoft Excel 2016 support
- Net Framework 4.6.1 Client Framework instead of .Net Framework 4.0 is now mandatory to execute the software
- Changed installation behavior
- MagicMaX SW 2.2.6 now automatically removes the previous MagicMaX SW
- MagicMaX SW 2.2.6 now automatically removes the previous MagicMaX SW
- Misc
- Improved performance when logging measurements
- Improved presentation of low dose rate values
- Improved compatibility when exporting measurements to MS Excel