Question
How to resolve the error “Data cannot be analyzed due to insufficient measurement time” in myQA machines FFF Profile analysis for 20 x 20c m² field size?
Answer
This error is noted with devices like MatriXX Evolution, MatriXX Resolution, MatriXX FFF, and StarTrack while connected to myQA Machines during direct measurement for FFF profiles with field sizes greater than 19 x 19cm2:
As a workaround, please use field sizes less than 20 x 20 cm2 in profile analysis for all FFF energies.
A solution is implemented for myQA 2024-001 and newer to resolve the restrictions with large field sizes in FFF profile analysis.