Question
How to update/replace the LFC license for myQA iON?
Answer
This is how to update/replace the LFC license for myQA iON:
Step 1
Stop services LFC-CS and all LFC-LS-[MachineName] services.
Step 2
Go to C:\Program Files (x86)\IBA Dosimetry\IBA Dosimetry LFC Service\converter\root\FileExchanges\forLFC and Save/Replace new LFCLicenses.xml
Step 3
Go to C:\Program Files (x86)\IBA Dosimetry\IBA Dosimetry LFC Service\converter\root\LogFiles and delete all folders/files.
Attention: This will delete all recorded logfiles for myQA iON for treatments in the past.
Step 4
Go to C:\Program Files (x86)\IBA Dosimetry\IBA Dosimetry LFC Service\converter and execute LFC-ServiceInstaller.exe as admin.
You can check that the new license was applied correctly when you go to C:\Program Files (x86)\IBA Dosimetry\IBA Dosimetry LFC Service\converter\LFC-LS\[Machine Name] and check if admin.xml was updated.
Step 5
Start services LFC-CS and all LFC-LS-[MachineName] services.